Table 2

Mean Ct/Cp values and %CV values for the undiluted OptiQual Enterovirus Run Control in different real-time RT-PCR settings

Extraction Method

Platform

Mean Ct/Cp

(%CV)


Primer-probe set 1

Primer-probe set 2


QiaAmp

7700 SDS

38.47 ± 3.94

(10.2%)

37.74 ± 1.35

(3.6%)


LC 480

33.69 ± 2.95

(8.8%)

34.48 ± 1.67

(4.8%)


EasyMag

7700 SDS

38.23 ± 3.88

(10.2%)

36.29 ± 0.91

(2.5%)


LC 480

32.81 ± 0.25

(0.8%)

32.78 ± 0.20

(0.6%)


Piqueur et al. Virology Journal 2009 6:95   doi:10.1186/1743-422X-6-95