Development of cost-effective real-time PCR test: to detect a wide range of HBV DNA concentrations in the western amazon region of Brazil

Abstract

Background

Currently there is a significant risk of infection with hepatitis B virus (HBV) during blood transfusion in high epidemic area. This is due to the pre-seroconversion window period, immunovariant viral strains and the presence of occult HBV infection (OBI). The aim of this study was to develop an in-house real-time PCR-based method, which was both ultra-sensitive and efficient offering an alternative method for nucleic acid testing (NAT).

Methods

A precore fragment with 109 bp was cloned and serial diluted to standard curve construction. The calibration of the HBV - DNA values was performed against OptiQuant® HBV-DNA Quantification Panel, Acrometrix Europe B.V.).

Results

From our in-house plasmid we prepared serial dilutions ranging from 2 × 103 – 2 × 109 copies/ml. The threshold was adjusted automatically during analysis and the data collected were analyzed by linear regression (r2 = 0.99). The limit of detection for the assay with pHBVRO standards was 2000/ml in a total reaction volume of 30 μl. We found a strong correlation between the two methods (r2 = 0.9965 and p < 0.0001). The regression line give us the following equation: Log 10 (IU/mL) = 0.9038Log 10 (copies/mL) − 1.0643, suggesting that 1 IU/mL = 15 copies/mL.

Conclusions

Therefore, we can affirm that the qHBVRO PCR can detect HBV DNA in individuals with hepatitis B at any stage of the disease showing high capacity for NAT screening in hepatitis b donors. This results of sensitivity could provide an advance for automation in blood banks and increasing safety of patients who receive blood transfusions.

Introduction

The hepatitis B virus (HBV) is one of the most common human pathogens and can cause hepatitis and aggressive and advanced liver disease, including cirrhosis and hepatocellular carcinoma [1]. Despite the availability of a vaccine, the implementation of preventive measures and serological screening in blood banks remains a major public health problem worldwide [2]. HBV can be transmitted perinatally, percutaneously, sexually or by horizontal transmission, especially among children, presumably through open cuts or sores [3].

Early detection of HBV surface antigen (HBsAg) significantly reduces the risk of infection through blood transfusions [4]. However, there are two situations in the course of infection where this early detection is currently ineffective: First, during the acute phase of infection, there is a window period where HBsAg may be undetectable in serum [5]. In another situation, occult infection which is defined as the presence of HBV DNA in the liver (with or without detectable serum HBV DNA) may be present during the persistent of infection in subjects who test negative for hepatitis B surface antigen (HBsAg). These subjects often have very low viral load (< 200UI/ml) [6]. Reducing the risk of transmission in these situations will require increased sensitivity the detection of HBV surface antigen (HBsAg), screening for antibodies to HBV core antigen (anti-HBc), and continued testing and implementation of NATs [7, 8].

Real-time polymerase chain reaction (qRT-PCR) has enabled the development of improved diagnostic tests offering greater speed while maintaining excellent levels of sensitivity and specificity [911]. qRT-PCR-based detection methods have been developed for the diagnosis of HBV and other pathologies in clinical laboratories [1214].

To successfully monitor viral load, it is important to diagnose viral replication, establish the prognosis of liver disease, to assess the risk of disease progression, to identify patients who need antiviral therapy and to monitor the virologic response to treatment. Currently there are several types of detection and quantitation assays in use, with varying levels of success [15, 16]. The aim of this study was to develop an in-house real-time PCR-based method, which was both ultra-sensitive and effective, offering a new NAT alternative.

Materials and methods

Clinical samples

This study included 134 patients with chronic HBV infection who were treated at the Viral Hepatitis Clinic Specialized Center of Research in Tropical Medicine in Rondonia (CEPEM). A control group of 30 donors, who all tested negative por ELISA for human immunodeficiency virus (HIV) 1 and 2, HBsAg, anti-HBc and anti-HCV, and who attended the blood bank of the State of Rondonia (FHEMERON) was included in the study. We also included 10 and 26 serum samples from individuals with chronic HCV and co-infection with HBV/HDV respectively.

Ethical consent

This study was approved by the Brazilian Institutional Ethics Committee of the Centro de Pesquisa em Medicina Tropical (CEPEM), with process number 107/10. Written, informed consent was obtained from each patient for the publication of this manuscript and any accompanying images.

DNA extraction

Viral DNA extraction was performed using the QIAamp DNA Mini Kit (Qiagen, Hilden, Germany) and 200 μl of serum according to the manufacturer’s instructions. Three samples with viral load known were tested: first with high viral load and the others medium and low viral load. After this were diluted with final volume 200 ul, 100 ul and 50 ul. Besides, HBV DNA was extracted from 22 samples of individuals with same profile serological: total isolated anti-HBc. These samples were diluted in 50 uL and 200 uL to optimize the final volume of extraction for samples with low viral load. Subsequently the samples were submitted to reaction for sensitivity analysis. Precipitated DNA was resuspended in elution buffer and stored at −20°C until further use. To avoid false-positive results, we followed strict procedures for nucleic acid amplification [17].

In-house testing

Primer concentrations were optimized using a concentration gradient ranging from 100–900 nM and SYBR® Universal PCR Master Mix (Applied Biosystems, Foster City, CA, USA). TaqMan® probe concentrations were similarly optimized using a concentration gradient ranging from 50–300 nM.

Ultra-sensitive real-time PCR

The assay was performed on an ABI 7500 platform (Applied Biosystems) with 30 μl reaction volumes containing 15 μl TaqMan® Universal Master Mix (Applied Biosystems), 3 μl HBVRO1 forward primer (5′-AGGAGGCTGTAGGCATAAATTGG 3′), 3 μl reverse primer (5′-GCACAGCTTGGAGGCTTG-3′), 0.6 μl probe (5′-FAM TCACCTCTGCCTAATC-3′-MGB, 6 μl extracted DNA and 2.4 μl of water.

Construction of the standard curve

To construct the standard curve, we initially used conventional PCR with amplification of a 109 bp fragment in the pre-core region according Kavita 2006 adapted, selecting five samples with known viral load. Approximately 50 ng of DNA was used per reaction with a final volume of 50 μl. Amplification was performed on an ABI Prism 7500 Veriti (Applied Biosystems) with an initial denaturation temperature of 94°C for 5 min, followed by 40 cycles of 94°C for 1 min, 58°C for 45 sec, 72°C for 1 min and a final extension of 10 min at 72°C. The selected fragment was purified, ligated to the p-GEM-T Easy® vector (Promega, New York, USA), cloned into a prokaryotic system and subsequent linearization with Pst I (Invitrogen™ Life Technologies, Carlsbad, CA, USA). Absolute quantitation was used to determine the exact number of DNA molecules for estimating viral load.

Inter- and intra-assay variation and reproducibility of real-time PCR

To determine intra-experimental variation, we tested the reproducibility of six HBV-positive sera with different viral loads, in duplicate, in the same reaction setup. The same set of samples was used in three experiments performed on different days, to estimate inter-experimental variation in the estimation of viral load. Reproducibility was estimated by calculating the coefficient of variation (CV), which is calculated as the ratio of the standard deviation and the mean of the replicates.

HBV DNA quantification

Calibration curves for HBV DNA were constructed using the OptiQuant® HBV-DNA Quantification Panel, (AcroMetrix Europe BV). Specifically, a serial dilution was prepared from the standards included in the kit, ranging from 2 × 102 – 2 × 106 IU/ml. For our in-house plasmid, pHBVRO, a serial dilution was prepared with a range 2 × 103 – 2 × 109 copies/ml. The concentration was measured spectrophotometrically both by using a NanoDrop® ND-1000 (Thermo Scientific NanoDrop Products, Wilmington, Delaware), and the measurements were recorded in units of nanograms per microliter, which was converted into copies per microliter by using the following equation: ([x ng/μL × 10-9] / [p-GEM-T Easy® vector and 109 HBV DNAbps × 660]) × 6.022e23 = y copies/μL. Using linear regression a standard curve was constructed, which was used to convert copies/ml to standard international units (IU/ml).

Quantification panel

To compare the performance of our in-house method (qHBVRO) with that of the commercial kit, OptiQuant® HBV-DNA Quantification Panel (AcroMetrix® Europe BV), we tested 100 serum samples collected from patients chronically infected with HBV.

Statistical analysis

The correlation between the AcroMetrix® test kit and the qHBVRO assay was calculated using GraphPad 5.0 (GraphPad software) and a two-tailed Pearson′s correlation test with a confidence interval of 95%. The units for measuring viral load (copies/ml and IU/ml) were transformed to log base 10.

Analytical specificity

We tested 30 samples from blood donors, 10 serum samples from mono-infected HCV patients, 28 samples from patients co-infected with HBV/HDV and 15 samples that were judged indeterminate for HBV surface antigen by the Serology Laboratory of Viral Hepatitis Clinic - IPEPATRO. All samples were submitted to qHBV PCR to determine viral load.

Analytical sensitivity

We selected 15 sera, which were considered indeterminate for HBsAg by ELISA when tested in duplicate and which had absorbance values within the gray zone, or ± 10% cut-off confidence interval. These samples were subjected to three separate assays, with each sample performed in duplicate to evaluate the performance of our assay in detecting uncertain samples.

Results

Analytical sensitivity and efficiency

From our in-house plasmid we prepared serial dilutions ranging from 2 × 103 – 2 × 109 copies/ml. The concentration of primers used was 300 nM, for both primers, and 100 nM for the probe. The threshold was adjusted automatically during analysis and the data collected were analyzed by linear regression (r2 = 0.99), Figure 1. The limit of detection for the assay with pHBVRO standards was 2000/ml in a total reaction volume of 30 μl. Two positive samples with known viral load were used as internal controls. HBV DNA extraction was patterned on elution of 200 μl because there was no significant variation in sensitivity of method in different volumes of patients with intermediate and high viral load. But the 22 samples of patients with serological profile “anti-HBc total isolated” eluted in 50 and 200 uL, 15 samples were negative on both elution, 5 samples were positive only in elution of 50 uL within which 2 samples were positive in both elution.

Figure 1
figure 1

qHBVRO standard curve as determined by linear regression.

Inter- and intra-assay variation and reproducibility of real-time PCR

The six HVB-DNA positive sera tested showed no statistical differences between repeats. The coefficient of variation was similar in both high and low viral load samples (0.01-0.16%), indicating the same efficiency of amplification for varying viral loads. There was no statistical difference in intra-and inter-assay variation (CV), which confirmed the reproducibility of the assay (Tables 1 and 2).

Table 1 Intra-experimental variation in the qHBVRO assay for serum samples
Table 2 Inter-experimental variation in the qHBVRO assay for serum samples

Validation of the qHBVRO method

Using linear regression, we found a strong correlation between the qHBVRO assay and the AcroMetrix® HBV-DNA kit (r2 = 0.998 and p <0.0001) as shown in Figure 2. Viral load values between the AcroMetrix® HBV-DNA and the qHBVRO assay were compared for 134 patients by Pearson’s correlation. We found a strong correlation between the two methods (r2 = 0.9965 and p < 0.0001). The regression line give us the following equation: Log 10 (IU/mL) = 0.9038Log 10 (copies/mL) − 1.0643, suggesting that 1 IU/mL = 15 copies/mL (Figure 3).

Figure 2
figure 2

WHO 1st International Standard (HBV ® kit Acrometrix DNA) as determined by linear regression.

Figure 3
figure 3

Analytical sensitivity of qPCR HBV demonstrated with 97 samples: 91 samples positive for HBsAg and 6 samples total anti-HBc isolated. These 6 samples were DNA-HBV positive detected within 50 clinical samples tested in patients with profile total anti-HBc isolated. The dashed line is the limit of analytical detection of qPCR HBV.

Analytical specificity and detection performance of indeterminate for ELISA samples and anti-HBc isolated samples

Of the 15 indeterminate for ELISA samples tested, six were positive for HBV DNA by both methods (Table 3). All samples that were classified as negative by ELISA were confirmed as such as shown. Of the 22 samples isolated anti-HBc eluted in 50 uL, 5 were positive (Figure 4).

Table 3 Data summary for indeterminate samples
Figure 4
figure 4

Analytical specificity of samples that tested positive, undetermined, negative and anti-HBc alone by ELISA using the qPCR HBV in house .

Discussion

Currently, there is a significant risk of infection with hepatitis B virus (HBV) during blood transfusions. This is due to the pre-seroconversion window period, immunovariant viral strains and the presence of occult HBV infection (OBI) [8, 18, 19]. Combined detection of HBsAg and anti-HBc constitutes an important strategy in donor screening which excludes the vast majority of OBIs [2022]. However, in many countries, especially in areas of low HBV prevalence (< 3%) these strategies are inefficient especially during the pre-seroconversion window period. However, in countries with high prevalence of positivity for anti-HBc, using this marker could involve both the exclusion of several potential donors leading to a decrease of blood products in various regions - such as the failure to block potential donors with OBI - keeping present the risk of post-transfusion HBV [18, 23]. The qHBVRO test presents an excellent alternative to HBV detection and quantification because it allows significant reduction in the risk of transmission during the window period as well as improved detection of occult HBV.

In this study we developed a qRT-PCR assay for the identification and quantification of HBV DNA with an efficiency of 94.06% and good correlation with the currently used commercial test: WHO 1st International Standard (HBV ® kit Acrometrix DNA), r = 0.998, p < 0.0001. The qHBVRO PCR test was 100 fold more sensitive, allowing detection of up to two thousand copies per ml of serum in HBV-infected individuals. Our method also proved to be more sensitive than other in-house qRT-PCRs [2426], and could detect occult hepatitis B infections as well as cases which were inconclusive by ELISA, using only 6 μl of DNA extracted from 200 μl of serum in final reaction volume of 30 μl for qHBVRO. However, was observed that in cases of samples with low viral load is important to consider a smaller volume in the elution of DNA. It is the case of 22 samples tested from individuals with isolated anti-HBc where 5 were positive in 50 uL elution and only 2 were positive on PCR qHBV. These results reinforce the importance that in case of occult infection or immunological window period, the concentration of DNA is an important factor to consider.

The high analytical specificity of the test, using samples from individuals that tested positive, negative and indeterminate for HBsAg by ELISA demonstrates that qHBVRO PCR can detect HBV DNA in individuals with hepatitis B at any stage of the disease, qualifying it as an important alternative NAT. The qHBVRO assay is highly reproducible, with low intra- and inter-experimental variation of between 0-1% (CV), whereas other in-house tests, which are considered to have good reproducibility, show experimental variation of 4.94–10.59% [2731].

Our method proved to be efficient, sensitive, specific and reproducible in the detection of occult HBV, and could therefore be used for nucleic acid testing (NAT) in blood banks to prevent HBV transmission by blood transfusion. The advantages of NAT relating to cost and effectiveness compared with serological diagnostics have been widely debated [2830]. It has been suggested that NAT offers advantages in many instances including occult infections, in the confirmation of viremia, for screening blood- and organ-donors, discriminating between patients with chronic or acute infection which has been resolved, diagnosis of perinatal infection, solving indeterminate serological results, monitoring patients on antiviral therapy and to identify the virus in immunocompromised individuals [3133].

Conclusion

In conclusion, the real-time PCR assay qHBVRO is appropriate for the quantification of HBV DNA in serum samples. This test is reproducible and proved be sensitive detecting samples with low viral load. Therefore, we can affirm that the qHBVRO PCR can detect HBV DNA in individuals with hepatitis B at any phase of disease showing good NAT screening for hepatitis B. Samples of patients anti-HBc positive isolated were selected and submitted to qHBVRO test to enhancing sensitivity the this results. This developed test may be automated and used in blood banks, increasing safety of patients who receive blood transfusions.

References

  1. Ganem D, Prince AM: “Hepatitis B virus infection–natural history and clinical consequences”. N Engl J Med 2004, 350: 1118-1129. 10.1056/NEJMra031087

    Article  PubMed  CAS  Google Scholar 

  2. Kitab B, Essaid EL Feydi A, Afifi R, Trepo C, Benazzouz M, Essamri W, Zoulim F, Chemin I, Alj HS, et al.: “Variability in the precore and core promoter regions of HBV strais in Morocco: characterization and impact on liver diasease progression. PLoS One 2012, 7: e42891. 10.1371/journal.pone.0042891

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  3. Bauer T, Sprinzl M, Protzer U: Immune control of hepatitis B virus. Dig Dis 2011, 29: 423-433.

    PubMed  Google Scholar 

  4. Kuhns MC, Kleinman SH, McNamara AL, Rawal B, Glynn S, Busch MP: “Lack of correlation between HBsAg and HBV DNA levels in blood donors who test positive for HBsAg and anti-HBc: implications for future HBV screening policy”. Transfusion 2004, 44: 1332-1339. 10.1111/j.1537-2995.2004.04055.x

    Article  PubMed  CAS  Google Scholar 

  5. Datta S, Banerjee A, Chandra PK, Chakraborty S, Basu SK, Chakravarty R: “Detection of a premature stop codon in the surface gene of hepatitis B virus from an HBsAg and antiHBc negative blood donor”. J Clin Virol 2007, 40: 255-258. 10.1016/j.jcv.2007.08.003

    Article  PubMed  CAS  Google Scholar 

  6. Raimondo G, Navarra G, Mondello S, Costantino L, Colloredo G, Cucinotta E, Di Vita G, Scisca C, Squadrito G, et al.: “Occult hepatitis B virus in liver tissue of individuals without hepatic disease”. J Hepatol 2008, 48: 743-746. 10.1016/j.jhep.2008.01.023

    Article  PubMed  Google Scholar 

  7. Saldanha J: Validation and standardisation of nucleic acid amplification technology (NAT) assays for the detection of viral contamination of blood and blood products. J Clin Virol 2001, 20: 7-13. 10.1016/S1386-6532(00)00149-9

    Article  PubMed  CAS  Google Scholar 

  8. Candotti D, Allain JP: “Transfusion-transmitted hepatitis B virus infection”. J Hepatol 2009, 51: 798-809. 10.1016/j.jhep.2009.05.020

    Article  PubMed  Google Scholar 

  9. Foy CA, Parkes HC: “Emerging homogeneous DNA-based technologies in the clinical laboratory”. Clin Chem 2001, 47: 990-1000.

    PubMed  CAS  Google Scholar 

  10. Arya SC, Agarwal N: “Hepatitis B vaccination of health care workers in Saudi Arabia”. Am J Infect Control 2005, 33: 613-614. author reply 612–613 10.1016/j.ajic.2005.05.022

    Article  PubMed  Google Scholar 

  11. Gunson RN, Abraham E, Carman WF: “Contamination with PCR detectable virus in a virus isolation quality assurance panel”. J Virol Methods 2006, 137: 150-151. 10.1016/j.jviromet.2006.05.018

    Article  PubMed  CAS  Google Scholar 

  12. Pas SD, Fries E, De Man RA, Osterhaus AD, Niesters HG: Development of a quantitative real-time detection assay for hepatitis B virus DNA and comparison with two commercial assays. J Clin Microbiol 2000, 38: 2897-2901.

    PubMed  CAS  PubMed Central  Google Scholar 

  13. Weiss MJ: “Beware! Uncle Sam has your DNA: legal fallout from its use and misuse in the U.S”. Ethics Inf Technol 2004, 6: 55-63.

    Article  PubMed  Google Scholar 

  14. Ronsin C, Pillet A, Bali C, Denoyel GA: “Evaluation of the COBAS AmpliPrep-total nucleic acid isolation-COBAS TaqMan hepatitis B virus (HBV) quantitative test and comparison to the VERSANT HBV DNA 3.0 assay”. J Clin Microbiol 2006, 44: 1390-1399. 10.1128/JCM.44.4.1390-1399.2006

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  15. Berger A, Preiser W, Doerr HW: “The role of viral load determination for the management of human immunodeficiency virus, hepatitis B virus and hepatitis C virus infection”. J Clin Virol 2006, 20: 23-30.

    Article  Google Scholar 

  16. Mackay IM, Arden KE, Nitsche A: “Real-time PCR in virology”. Nucleic Acids Res 2002, 30: 1292-1305. 10.1093/nar/30.6.1292

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  17. Kwok S, Higuchi R: Avoiding false positives with PCR. Nature 1989, 339: 237-238. 10.1038/339237a0

    Article  PubMed  CAS  Google Scholar 

  18. Liu CJ, Kao JH, Chen DS: “Kinetics of hepatitis B virus reactivation after chemotherapy: more questions than answers”. Gastroenterology 2006, 131: 1656-1657.

    Article  PubMed  Google Scholar 

  19. Panigrahi R, Majumder S, Gooptu M, Biswas A, Datta S, Chandra PK, Banerjee A, Chakrabarti S, Bandopadhyay D, et al.: “Occult HBV infection among anti-HBc positive HIV-infected patients in apex referral centre, Eastern India”. Ann Hepatol 2012, 11: 870-875.

    PubMed  CAS  Google Scholar 

  20. Allain JP: “Occult hepatitis B virus infection: implications in transfusion”. Vox Sang 2004, 86: 83-91. 10.1111/j.0042-9007.2004.00406.x

    Article  PubMed  Google Scholar 

  21. Fang CT: “Blood screening for HBV DNA”. J Clin Virol 2006,36(1):30-32.

    Article  Google Scholar 

  22. Keyvani H, Agah S, Kabir A, Alavian SM: “Prevalence and risk factors of isolated anti-HBc antibody and occult hepatitis B infection in hemodialysis patients: a nationwide study”. Ann Hepatol 2013, 12: 213-219.

    PubMed  Google Scholar 

  23. Chakrabarti S, Bandopadhyay D: “Occult HBV infection among anti-HBc positive HIV-infected patients in apex referral centre, Eastern India”. Ann Hepatol 2012, 11: 870-875.

    PubMed  Google Scholar 

  24. Paraskevis D, Beloukas A, Haida C, Katsoulidou A, Moschidis Z, Hatzitheodorou H, Varaklioti A, Sypsa V, Hatzakis A: Development of a new ultra sensitive real-time PCR assay (ultra sensitive RTQ-PCR) for the quantification of HBV-DNA. Virol J 2010, 7: 57. 10.1186/1743-422X-7-57

    Article  PubMed  PubMed Central  Google Scholar 

  25. Paraskevis D, Haida C, Tassopoulos N, Raptopoulou M, Tsantoulas D, Papachristou H, Sypsa V, Hatzakis A: Development and assessment of a novel real-time PCR assay for quantitation of HBV DNA. J Virol Methods 2012, 103: 201-212.

    Article  Google Scholar 

  26. Lole KS, Arankalle VA: “Quantitation of hepatitis B virus DNA by real-time PCR using internal amplification control and dual TaqMan MGB probes”. J Virol Methods 2006, 135: 83-90. 10.1016/j.jviromet.2006.02.004

    Article  PubMed  CAS  Google Scholar 

  27. Pawlotsky JM, Bastie A, Hezode C, Lonjon I, Darthuy F, Remire J, Dhumeaux D: “Routine detection and quantification of hepatitis B virus DNA in clinical laboratories: performance of three commercial assays”. J Virol Methods 2000, 85: 11-21. 10.1016/S0166-0934(99)00149-4

    Article  PubMed  CAS  Google Scholar 

  28. Noborg U, Gusdal A, Pisa EK, Hedrum A, Lindh M: “Automated quantitative analysis of hepatitis B virus DNA by using the Cobas Amplicor HBV monitor test”. J Clin Microbiol 1999, 37: 2793-2797.

    PubMed  CAS  PubMed Central  Google Scholar 

  29. Yoshikawa A, Gotanda Y, Itabashi M, Minegishi K, Kanemitsu K, Nishioka K: HBV NAT positive [corrected] blood donors in the early and late stages of HBV infection: analyses of the window period and kinetics of HBV DNA. Vox Sang 2005, 88: 77-86. 10.1111/j.1423-0410.2005.00602.x

    Article  PubMed  CAS  Google Scholar 

  30. Leung VK, Lee CK, Chau TN, Cheung WI, Lo FH, Lai KB, Lin CK: “A probable case of transfusion-transmitted hepatitis B virus infection in an immunosuppressed recipient caused by an occult HBV-infected donor with negative ID-NAT”. Transfus Med 2010, 20: 276-277. 10.1111/j.1365-3148.2010.00997.x

    Article  PubMed  CAS  Google Scholar 

  31. Tani Y, Aso H, Matsukura H, Tadokoro K, Tamori A, Nishiguchi S, Yoshizawa H, Shibata H: Significant background rates of HBV and HCV infections in patients and risks of blood transfusion from donors with low anti-HBc titres or high anti-HBc titres with high anti-HBs titres in Japan: a prospective, individual NAT study of transfusion-transmitted HBV, HCV and HIV infections. Vox Sang 2012, 102: 285-293. 10.1111/j.1423-0410.2011.01561.x

    Article  PubMed  CAS  Google Scholar 

  32. Chatterjee K: An improved nat assay with enhanced sensitivity for hbv for donor blood screening. J Lab Physicians 2012, 4: 127.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Liu CJ, Chen DS, Chen PJ: “Epidemiology of HBV infection in Asian blood donors: emphasis on occult HBV infection and the role of NAT. J Clin Virol 2006, 36: 33-44.

    Article  Google Scholar 

Download references

Acknowledgments

This work was supported by IPEPATRO, CEPEM, Universidade Federal de Rondônia and Fundação Oswaldo Cruz- Rondônia. Our thanks to Larissa Deadame F Nicolete by collaboration in the statistical analyzes.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Alcione de Oliveira dos Santos.

Additional information

Competing interests

The authors declare that they have no financial or competing interest with this article.

Authors’ contributions

AOS participated in the design of the study, drafted the manuscript and in its design and coordination. LFBS participated in the PCR amplification and sequencing process. DSV participated in the design of the study. LMB participated in the elaboration of the manuscript. JMS participated in the design of the study. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited.

Reprints and permissions

About this article

Cite this article

de Oliveira dos Santos, A., Souza, L.F.B., Borzacov, L.M. et al. Development of cost-effective real-time PCR test: to detect a wide range of HBV DNA concentrations in the western amazon region of Brazil. Virol J 11, 16 (2014). https://doi.org/10.1186/1743-422X-11-16

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1743-422X-11-16

Keywords